TA300 - Temperature Transmitter for High Temperature
  • Measuring range from -50℃...200℃ to -200℃...600℃
  • Compact design
  • Outputs programmable
  • Process connections selectable
  • Stainless steel housing
Applications
  • Circulating water/cooling liquids
  • Hydraulics and pneumatics
  • Oil and Gas
  • Machinery manufacturing
  • Water treatment
  • Construction automation
Specifications
Power supply
    Current output model 10…30Vdc
    Voltage output model 14…30Vdc
Current consumption
    2-wire Same as current output ( 2-wire 4…20mA)
    3-wire ≤20mA
2-wire current output
    Output 2-wire 4.20mA
    Load RA(Ω) RA≤(Us-10)V/0.02A
    Linearity ≤±0.2% of F.S.
3-wire current output
    Output 3-wire 4...20mA,0...20mA
    Load RA(Ω) RA≤(Us-3)V/0.02A
    Linearity ≤±0.2% of F.S.
Voltage analog output
    Output 3-wire 0...5V,1...5V, 0...10V
    Load RA(Ω) RA≥5K
    Linearity ≤±0.2% of F.S.
Sensor PT1000 Class A
Accuracy ≤±0.25% of F.S.
Stability(drift/year) ≤±0.1% of F.S.
Temperature
    Medium(measuring range) -50…120℃; -50…200℃; -50…400℃; -50…600℃ (measuring range customizable)
    Ambient -40 +85℃
Material
    Housing 304 stainless steel
    Wetted parts 316 stainless steel
Protection class IP65(solenoid plug); IP67(M12×1plug)
Electrical connection M12x1plug; solenoid plug